LWIR camera, microbolometer sensor, 320 x 240 pixels

The Pearleye P-007 LWIR camera incorporates an uncooled microbolometer sensor with 320 x 240 pixels resolution. With its maintenance-free sensor, a temperature reference element, and a Peltier temperature stabilization, the camera reliably detects temperature differences <80 mK. Built-in image correction features ensure an excellent image quality.

  • Amorphous silicon uncooled microbolometer focal plane array (FPA),
    320 x 240 pixels, sensor time constant 7 ms
  • 35 μm x 35 μm cell size, effective chip size 11.2 x 8.4 mm 
  • Spectral response: 8 - 14 μm (LWIR)
  • NETD ≤ 80 mK@ 303 K @ f/1.0
  • Temperature measurement range: -20°C to +80°C @ f/1.0
  • Temperature reference element and Peltier temperature stabilizing
  • Frame rate 40 fps (40 Hz)
  • Built-in electromechanical calibration shutter
  • Preprocessing functions included
  • Including 18 mm lens, f/1.0 (field of view: 34.6° x 26.3°)
  • Options
    • 12 mm lens, f/0.85
    • 35 mm lens, f/1.0
    • Other lenses on request

Models:
Pearleye P-007 LWIR, temperature monitoring range -20°C to  +80°C     
Pearleye P-007 LWIR High Temp, temperature monitoring range 0°C to +200°C

Pearleye P-007 LWIR
Interface IEEE 802.3 1000baseT
Resolution 320 x 240
Spectral range LWIR, 8 - 14 µm
Sensor ULIS UL 03 08 1
Sensor type Microbolometer
Sensor size No standard size
Cell size 35 µm x 35 µm
Lens mount M65 x 0.5
Max frame rate at full resolution 40 fps
Temperature measurement -20 °C ... +80 °C, High Temp Version: 0 °C ... +200 °C
NETD ≤ 80 mK@ 303 K @ f/1.0
A/D 14 bit
Output
Bit depth 12 bit
Mono modes Mono12
Operating conditions/Dimensions
Operating temperature 0°... +35 °C
Power requirements (DC) 12 V
Power consumption (12 V) 18 W
Mass 830 g
Body Dimensions
(L x W x H in mm)
133.7 x 90 x 86 mm, incl. lens and connectors
Regulations CE, RoHS (2002/95/EC)

Download Technical drawing (click here)

  • Shipped with various built-in correction data sets
  • Factory adjusted bad pixel correction
  • Background (FPN) correction
  • Gain/offset correction (NUC/non-uniformity correction) for each pixel
  • Drift compensation
  • Temperature linearization (LUT)
  • Continuous mode (image acquisition with maximum frame rate)

In combination with AVT’s AcquireControl software, extensive image analysis functions are available:

  • Pseudo color LUT with several color profiles
  • Auto contrast
  • Auto brightness
  • Temperature measurement
  • Analyze multiple regions (rectangular, circle) within the image
  • Real-time statistics and histogram display
  • … and more

The Pearleye P-007 LWIR is a maintenance-free, robust, compact LWIR camera with excellent image quality and precise temperature measurement. It detects subtle temperature differences with high precision.

  • OEM Applications
  • Surveillance
  • Automation
  • Quality control
  • Science and research